  • Cleaning method for oil cup of insulation oil dielectric loss tester

    The insulation oil dielectric loss tester Việc sử dụng nguồn điện siêu cao áp có thể giúp nhiều công nhân điện lực thực hiện các thử nghiệm điện khác nhau một cách thuận tiện hơn.

    HTY_2068

    Cleaning the oil cup is a very important preparation work when measuring the loss value of insulating oil with an insulation oil dielectric loss tester. Some unreliable measurement results are often caused by incomplete cleaning of the oil cup. Therefore, strict cleaning methods must be followed to obtain reproducible and reliable measurement results.

    When conducting the identification test of insulation oil loss coefficient, the oil cup should be thoroughly cleaned before each test. The cleaning steps are as follows:

    a. Completely remove the oil cup and clean all components with chemically pure petroleum ether (distillation range 60-90 ℃) and benzene in sequence.

    b. Rinse all components with acetone, then clean with neutral detergent.

    C. Boil all components in 5% trisodium phosphate in distilled water for 5 minutes, then rinse several times with distilled water.

    d. Boil all components in distilled water for at least 1 hour.

    e. Dry all parts in an oven controlled at 105-110 ℃ for at least 1 hour.

    F. When all components have cooled down to a point where they won't be hot to the touch, assemble the oil cup.

    Note: In the general routine test of insulation oil loss coefficient, the cleaning method of the oil cup can be simplified by omitting items c and d of the above cleaning method, and rinsing all components with distilled water several times instead. Then proceed directly to option e.

    In addition, when conducting routine tests on a batch of oil samples continuously, if the loss factor of the previous oil sample is less than the specified value, it is not necessary to clean the oil cup during the next oil sample production. However, a second oil sample must be used and the oil cup must be rinsed three times or more.

    Bạn cũng có thể thích

    • Có bất kỳ câu hỏi nào!

      ĐỪNG ngần ngại LIÊN HỆ VỚI CHÚNG TÔI BẤT CỨ LÚC NÀO
      Liên hệ chúng tôi
    • Ghé thăm chúng tôi

      Tòa nhà 6, Khu công nghiệp Cơ khí Quang học Thung lũng Quang học, 84 Đường Gaoxin 5,
      Khu phát triển công nghệ mới Hồ Đông, thành phố Vũ Hán, tỉnh Hồ Bắc, Trung Quốc

    • Gửi email cho chúng tôi

      qiao@hvtest.cc

    • Gọi cho chúng tôi

      +86-15871365102